Air-O-Cell Sampling Guide


Zefon's Air-O-Cell is a linear spore trap with an impaction trace and is used for the collection of airborne bio-aerosols including fungi, pollen, and particulate.Spore traps are appropriate for air sampling at the beginning of an investigation to set a baseline and as a final check for clearance sampling. They can also be used for monitoring during the remediation process.

The Air-O-Cell cassette is attached to a Zefon Biopump Plus which is capable of drawing 15 litres per minute.The cassette is sent to Eurofins, where it is disassembled and the slide is stained and examined microscopically by a qualified analyst.

Sampling times depend upon the environment being sampled. The table below has suggested times for various environmental conditions.

Suggested Sampling Times
Environment Flow Rate Duration
 Outside 15 LPM  5 - 10 Minutes
 Clean Room 15 LPM 10 Minutes
 Indoor - Occupied 15 LPM 5 Minutes
 Basement or Dusty Room 15 LPM 3-5 Minutes
 Wall Cavity 15 LPM 1 Minute

 

Procedure

  • Check the expiry date of the Air-O-Cell is current and remove the seal from the bottom of the cassette and retain to reseal when sampling is complete.
  • Attach the cassette to a calibrated Bio-Pump.
  • Remove the seal from the top of the cassette and retain to reseal when sampling is complete.
  • Set the desired run time for sampling.
  • Press "start" and record the sampling time on the Chain of Custody form. The Bio-Pump will turn itself off.
  • Disconnect Air-O-Cell from the Bio-Pump and reseal cassette
  • Record sample information on cassette and Chain of Custody
  • Store cassette in a cool, dry area (<40oC).

There is no maximum time for the analysis to be completed but it is advised to have the count & identification determined within the shortest possible time.

Please refer to the following Zefon Training Videos.